2004 ICS Paris
740
Chung B¹, Lee T², Yang J H²
1. Dept of Urology,
2. Dept of Urology,
THE DIAGNOSTIC VALUE OF PORTABLE BLADDER VOLUME MEASUREMENT SYSTEM (BVMS) WITH REAL BLADDER IMAGE IN THE MEASUREMENT OF BLADDER VOLUME ACCORDING TO THE DIFFERENT ANGLING OF TRANSDUCER
Hypothesis / aims of study
Several portable bladder volume measurement systems (BVMS) are commercially available.
However, there are several shortcomings of such systems, which make the measurement of the bladder volume inaccurate. The most important shortcoming when using a mechanical system on the abdomen is the undetermined angulation of the transducer, because we cannot see the real-time images and revise the angulation immediately.
So we assessed the volume estimation of bladder with two angles (10 and 30 craniocaudal direction) of the transducer by on examiner and investigated the accuracy determined by estimated and catheterized bladder volume. And we investigated the effect of angulation on the accuracy by analyzing the real bladder images which contain the total bladder images completely or incompletely on the Field-of-View(FOV).
Study design, materials and methods
Sonographic studies of bladder volume by a newly developed portable (2.4kg) BVMS (BioCon-500,
Results
Prefereable agreement between BVMS estimations and catheterized volumes was found for both cranio-caudal angles with a slight underestimation, but better for 30 degree estimations (no significance). The 10 degree angled ultrasound estimation showed correlation with bladder volumes of 0.980 (R²=0.961) and a mean difference from the true bladder volumes of 8.35ml (95% confidence interval 4.93 to 11.78ml). The 30 degree angled ultrasound estimation showed correlation with bladder volumes of 0.982 (R²=0.965) and a mean difference from the true bladder volumes of 3.21ml(95% confidence interval 0 to 6.45ml). In cases with less than 200ml of true volume, there was no case with bladder images swerved from FOV for both cranio-caudal angles. In cases with more than 200ml of true volume, 28.6% and 50.0% of patients, examined on 10 and 30 degree angles respectively, showed bladder images swerved from FOV.
Interpretation of results
Volume estimation using this BVMS showed very good agreement with true volumes at both 10 and 30 degree craniocaudal angles of the transducer. At more than 250ml of bladder volume, volume estimation at 30 degree craniocaudal angles of the transducer rather than 10 degree is recommended because of FOV limitation on ultrasound.
Concluding message